Ammonia Monitor for Wastewater
Robust and Reliable Technology. Used to monitor the wastewater plant intake
Robust and Reliable Technology. Used to monitor the wastewater plant intake
The MS3500 is the first ammonia monitor for Raw Waste Water. Its rugged construction and design allows it to reliably measure Ammonia at the very front end of a waste water treatment plant, giving the operator a level of control and knowledge presently unknown to the industry. It is used to:
Ammonia levels at the intake can change a lot from day to day and from hour to hour. No instrument to date is able to measure ammonia at the raw wastewater inlet (soon after the screens) without an extensive maintenance regime or filtration system. If there was an instrument able to measure ammonia levels soon after the screens operators could:
Spec | Value |
---|---|
Dynamic Range | 1 ppm – 200 ppm in water |
Absolute Accuracy | Ammonia in Water: ±10% |
Repeatability | Repeated measurement, 5ppm Ammonia in Water: ±2% |
Sampling Frequency | 60/30 minutes (Contact us for other sampling periods) |
Relays | 2x, User Programmable function |
Analogue Output | 4 – 20 mA, Loop Resistance 40 Ω – 1k Ω |
Digital Outputs | Modbus, Profibus as options |
Typical Power Consumption | 70 W in normal operation |
NaOH Requirement | 10 l of 1 M NaOH per 15 Weeks |
Enclosure | GRP |
Operating Temperature | 1 °C – 50 °C |